Automatic Flash Sterilizer
                                
                             
                            
                                - Should be a table top autoclave for Ophthalmic Applications.
 
                                - Two automatic programmes approx. at 2.2 bar at 134 degrees C and 1.1 bar at 121 degree C. The equipment should have automatic pressure control switch / automatic water control device to ensure that the equipment does not run dry. 
 
                                - Should have flash cycle for rapid sterilization and should have an option for liquid cycle. 
 
                                - Should have Air Pump for closed door drying. 
 
                                - Should have rapid warm up facility. Built in reservoir to store water required to produce steam, and used water separately, for easy decantation. 
 
                                - The system should be equipped with required safety features. The door should have double locking safety feature and should open only with atmospheric pressure in the chamber. 
 
                                - Should have automatic cut-off to prevent overheating and cut-off for insufficient water, the machine should not start without sufficient water. 
 
                                - Should have a minimum chamber capacity of 20 litres or above. 
 
                                - Should have pressure display and temperature display. 
 
                                - Unit should function with 200-230Vac, 50/60 Hz input power supply.